Most Common Applications
Aerospace: Alignment, Tooling & Mold Certification, Part Inspection
Automotive: Tool Building & Certification, Alignment, Part Inspection
Metal Fabrication: OMI, First article inspection, Periodic Part Inspection
Molding/Tool & Die: Mold and Die Inspection, Prototype Part Scanning
FaroArm Test Methods – (Test methods are a subset of those given in the B89.4.22 standard.)
Single Point Articulation Performance Test (Max-Min)/2:
The probe of the FaroArm is placed within a conical socket, and individual points are measured from multiple approach directions. Each individual point measurement is analyzed as a range of deviations in X, Y, Z. This test is a method for determining articulating measurement machine repeatability.
Volumetric Maximum Deviation:
Determined by using traceable length artifacts, which are measured at various locations and orientations throughout the working volume of the FaroArm. This test is a method for determining articulating measurement machine accuracy.
Hardware Specifications
Operating Temp range: 10°C to 40°C (50°F to 104°F)
Temperature Rate: 3°C/5min. (5.4°F/5min. Max)
Operating Humidity Range: 0-95%, noncondensing
Power Supply: Universal worldwide voltage
85-245VAC, 50/60 Hz